ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test

Share:

ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Tester | UnitedTest

UnitedTest supplies ASTM F2345 cyclic fatigue testing machines for ceramic modular femoral heads on metallic taper cones for total hip arthroplasty (THA), supporting comparative mechanical fatigue evaluation for orthopedic hip implant R&D and quality verification.


ASTM F2345 specifies laboratory cyclic fatigue test procedures for ceramic modular femoral heads assembled onto metallic taper cones for total hip arthroplasty (THA). It provides comparative mechanical evaluation rather than direct prediction of in‑vivo clinical performance. UnitedTest manufactures reliable ASTM F2345 fatigue test equipment to help orthopedic implant labs assess cyclic fatigue strength, validate design, raw material quality and taper connection stability of ceramic femoral head prostheses.


Test principle:

Ceramic hip femoral heads suffer clinical risks of fracture under repeated physiological hip joint loads. This standard uses sinusoidal cyclic compressive loading in a physiological‑simulating liquid environment to accelerate fatigue damage at the head‑cone taper interface.

The test works as a comparative bench model: cyclic stress accumulates within the ceramic head; specimens either survive 10 million cycles or experience brittle fracture. Post‑fatigue static burst test quantifies residual strength for heads without visible fracture.

Important note: laboratory load magnitudes and boundary conditions differ from real human body loading; results compare relative performance of different designs/materials instead of predicting clinical service life directly.

ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test


The three specific test methods

On-Axis Fatigue Test Method
  • Load axis angle L = 0 ± 1° (load collinear with the femoral neck axis)

  • Head loaded either through a hardened metal 100 ± 1° cone or with the contact ring protected by a copper ring, in accordance with ISO 7206-10

  • Sinusoidal waveform from peak load down to a minimum of 10 % of the load magnitude

  • Cyclic loading continues to 10 million cycles without failure, or until fracture

  • Minimum of five specimens per test group.

Off-Axis Fatigue Test Method
  • Load axis angle L = 30 ± 1° (simulating a more adverse/off-centre joint reaction direction)

  • Uses a polymeric spherical concave component whose segment diameter is 0.643 × the head diameter; the segment diameter must not change during the test.

  • Same sinusoidal waveform (minimum = 10 % of peak) and same ≤30 Hz limit

Post-Fatigue Burst Test Method
  • Performed after the on-axis or off-axis fatigue test

  • Static burst test according to ISO 7206-10, in air at ambient conditions

  • Purpose: determine the residual static strength of heads that survived cyclic loading.


Test Specimen info: 

  • Test articles: ceramic modular femoral heads paired with metallic taper cones. Acceptable ceramics: high‑purity dense alumina (ASTM F603), alumina‑zirconia composite (ISO 6474‑2), Y‑TZP zirconia (ISO 13356).

  • Specimens can be actual hip prostheses or geometry‑equivalent simplified cone‑neck models. Metallic cones must match final implant material, dimensional tolerances and surface finish.

  • Nominal head sizes: typical 22 mm, 26 mm, 28 mm, 32 mm, 36 mm, 40 mm, 44 mm spherical diameters.

  • Sample quantity: minimum five specimens per test group.

  • Mandatory rule: brand‑new cones and heads for every single test. Specimens survived testing cannot be used clinically after laboratory loading.

  • Geometric parameters to characterise: bore angle, cone angle, circularity, taper interface length, surface roughness R_a / R_z (DIN 4768), major/minor bore diameters.


Test Equipment for ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test

Fatigue testing machine

Servo‑hydraulic load frame: rated up to 50 kN fatigue capacity; shall comply with ASTM E‑4 force‑verification requirements. 

Dynamic applied force shall stay within ± 2 % of maximum compressive load throughout testing.

Loading fixtures

capable of sustaining static fracture‑level forces; force vector must pass through femoral head geometric centre. 

Fragment‑shielding safety enclosure is required for destructive cyclic testing.

Temperature‑controlled bath

maintains physiological solution (e.g. Ringer’s solution) at 37 °C.

Room‑temperature testing is allowed only if validated not to affect results.

Auxiliary parts

hardened metal conical loading adapter, disposable copper contact rings (per‑test replacement), polymeric concave off‑axis loading insert.


Test parameters

ParameterValue / Requirement
Test mediumPhysiological solution at 37 °C
Max test frequency≤ 30 Hz
Fatigue waveformSinusoidal compression; minimum load = 10 % of peak compressive load magnitude
Total cycle limit10 000 000 cycles (stop‑point if no fracture)
On‑axis load‑axis angle0 ± 1°
Off‑axis load‑axis angle30 ± 1°
Installation seating force2 kN static pre‑load (laboratory standard for repeatable assembly)
Post‑fatigue burst test delayWithin 1‑2 days after fatigue cycles finish
Force accuracyDynamic load tolerance ± 2 % of peak compressive force


Test Procedures of ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test

  1. Sample Characterisation: Document head‑cone geometry, roughness and material properties, complying with ISO 7206-10 (clauses 6.1 and 6.3.. 

  2. Specimen Preparation: Clean heads and cones following laboratory cleaning protocol (Appendix X2 reference workflow: tap‑water rinse → 1 %‑detergent ultrasonic wash 15 min → multi‑step distilled‑water rinsing → air drying).

  3. Assembly: Assemble head onto metallic cone applying 2 kN static installation pre‑load according to ISO 7206‑10 (clause 7.2). Mount assembled specimen into load fixture.

  4. Environmental setup: Submerge test region in 37 °C physiological solution.

  5. Fatigue testing: Choose either on‑axis or off‑axis configuration, set frequency ≤ 30 Hz, apply sinusoidal cyclic compression until specimen fractures or reaches 10 million cycles. Record failure occurrence and completed‑cycle count. Inspect and replace damaged fixtures / copper rings after each specimen.

  6. Post‑fatigue burst test: Transfer non‑fractured specimens within reasonable time window and perform static burst test according to ISO 7206‑10 under ambient air.

  7. Reporting: Document manufacturer information, head size, ceramic material, cone material, taper geometry, failure definition, test‑method type, frequency, peak load, load‑axis angle, contact‑loading setup, cycle counts, pass/fail status, time interval between fatigue and burst test. Optional supplementary data includes taper dimension tolerances, surface roughness, sterilisation method of femoral heads.

ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test


Industrial‑Field Application

  • Orthopaedic / medical device industry — total hip replacement (THR) manufacturers and ceramic component suppliers (alumina, ZTA/alumina-matrix composite, Y-TZP zirconia).

  • Design verification & material selection — comparing head designs, cone/bore taper geometry, surface finish, materials, manufacturing routes and sterilization effects on cyclic load-carrying ability.

  • Regulatory submission — FDA-recognized consensus standard used to support 510(k)/PMA and FDA guidance for ceramic hip systems; also used for EU MDR technical documentation .

  • Contract test laboratories / CROs — e.g. labs offering "PI-14 static/dynamic compression" fatigue testing of ceramic head systems built to meet ASTM F2345 and FDA guidance .

  • Quality control, benchmarking and failure investigation — including assessing risks such as head fracture, head dislodgement from the cone, cone wear from loosening/rotation, and cone fracture (X1.3).


Related Standard: 

ASTM F1800

Standard practice for cyclic fatigue testing of metal tibial tray components of total knee joint replacements

ASTM F2083Specification for Knee Replacement Prosthesis, specifies reference load magnitude, specimen quantity and run‑out cycle counts for tibial tray testing.
ISO 7206-10

Determination of resistance to static load of modular femoral heads (used for the burst test)

ASTM F3210Fatigue testing for knee femoral components
ASTM F2777endurance test for polyethylene tibial inserts.
ASTM F2009axial disassembly (pull-off) force of taper connections of modular prostheses
ISO 7206-13torque/head-fixation (torsion) test of the head–stem taper 
ISO 11491impact resistance of ceramic femoral heads
ISO 7206-4 / ISO 7206-6

endurance of stemmed femoral components and of the neck region; ASTM F2580 for modular proximally fixed stems; 

ASTM F2068 is the US analogue of ISO 7206-4 

ASTM F1814Guide for evaluating modular hip and knee joint components
ASTM F3161FEA of metallic TKA femoral components under closing conditions; the companion simulation route, and a recommended way to pick the worst-case condyle/size.
ASTM F2345

Standard Test Methods for Determination of Cyclic Fatigue Strength of Ceramic Modular Femoral Heads

ASTM F1264Standard specification and test methods for intramedullary fixation devices


Keywords: ASTM F2345, ceramic modular femoral head fatigue tester, cyclic fatigue strength test, total hip arthroplasty implant test, THA femoral head testing machine, hip prosthesis fatigue test, orthopedic implant mechanical test, ceramic hip head taper cone test, medical implant fatigue testing equipment, modular hip joint test, bioceramic implant fatigue test, orthopedic device lab tester, hip implant R&D test machine, femoral stem taper connection test, medical device fatigue test rig

Related products and device

ASTM F2345 Ceramic Modular Femoral Head Hydraulic Servo fatigue dynamic testing machine

Hydraulic fatigue dynamic testing machine mainly used at a variety of material and components, parts for both dynamic and static mechanical property testing, include tension, compression, low cycling test, match with high-low temperature chamber can execute the temperature test.

ASTM F2345 Dual axis Torsion & Bend Fatigue Testing machine

Electronic Dynamic Universal Testing Machine/ Cyclic Testing Equipment widely used at material and components, parts dynamic mechanical property test, include tension, compression, low cycling test, match with high-low temperature chamber can execute the temperature test.

ASTM F2345 intramedullary fixation device tester

Electronic Fatigue Testing Machine designed for long-term cyclic load testing of metals, composites, and structural components, offering precise electronic control, repeatable results, and compliance with industrial standards.

ASTM F2345 ceramic hip head taper cone test

Spinal Implant Constructs nail screw static and fatigue testing machine used to check the fatigue failure of medical nail, Loads are typically applied in a constant-amplitude, load-controlled sinusoidal waveform that runs more than 5 million times.

ASTM F2345 Static bending universal testing machine

UnitedTest universal testing machine with various test fixture, like peel, flexure, puncture, tear, pneumatic grip, belt tension etc., can be used to almost all materials include but not limited to steel rod, rubber, steel wire, plastic, seat belt, textile, wood, panel etc., mechanical performance inspection.

4 points bending fixture

The four point bending flexural test provides values for the modulus of elasticity in bending, flexural stress ,flexural strain and the flexural stress-strain response of the material. This test is very similar to the three point bending flexural test.

3 points bending grip

3 points bending grip used to measure the flexural modulus or flexural strength of a plastic material, conform with stanard ASTM D790, ISO 178.

intramedullary fixation devices torsion tester

Torsion tester is mainly used to execute torsion test for various kinds materials, manual loading, high accuracy torsion transducer to measure torque, photoelectric encoder to measure torsion angle, digital display test result. Easy to operate and with budgetary price, most suitable for research institute and colleague, factory laboratory etc. to check the torsion character of different materials.

Related Standard

ISO 7206-10 Femoral Head Static Compression and Tensile test –

ISO 7206-10: Implants for surgery --Partial and total hip joint prostheses -- Part 10: Determination of resistance to static load of modular femoral heads. 


It specifies a method for determining the static load resistance of the connection between a modular femoral head and a tapered cone of a femoral stem (the "head/cone junction") in hip joint prostheses. Include static compression (fracture) and tensile (disassembly) tests to quantify the load capacity of modular femoral heads with conical taper (Morse taper) connections. 


ASTM F1717 Spinal Implant Constructs Test –

ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Modelis a fundamental standard for evaluating the static and fatigue mechanical performance of spinal implant assemblies. Unlike ISO 12189, which focuses on a single device(an interbody cage), ASTM F1717 tests the entire posterior spinal construct—typically rods, screws, and cross-connectors—stabilizing a segment where a vertebral body has been removed (a "vertebrectomy" or "corpectomy" model). It simulates worst-case spinal instability to assess the implant's ability to maintain alignment and support loads.

ASTM F382 Metallic Bone Plates fatigue bending test –

ASTM F382 Standard Specification and Test Method for Metallic Bone Plates

ASTM F382 unified terminology, geometric classification, material requirements, marking & packaging rules, and mechanical test protocols for metallic bone plates used in surgical skeletal internal‑fixation. It define two mandatory test methods:  Single‑Cycle Bend Test and Bending Fatigue Test. Both adopt four‑point bending configuration. It is of great significance for ensuring the safety and reliability of metallic bone plates, which are widely used to align and fix bone segments in orthopedic reconstructive surgeries. 

ASTM F1798 Spinal Implant Subassembly Static and Fatigue Testing System –

ASTM F1798 Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants.

ASTM F1798 provides standardized methods for mechanically testing the interconnections within spinal implant systems. It is for evaluating the uniaxial static/fatigue strength and loosening resistance of interconnection mechanisms in spinal arthrodesis implant subassemblies, providing a standardized way to characterize mechanical performance of connections like rod-clamp, screw-rod, and hook-rod assemblies. It is critical for design validation, regulatory compliance, and clinical safety in the spinal implant industry.

ASTM F2706 Fatigue Static Test of Spinal implant constructs –

ASTM F2706 is a critical biomechanical evaluation tool in the medical device industry, specifically for spinal implants. ASTM F2706 establishes standardized mechanical test methods to evaluate the static (strength) and fatigue (long-term durability) performance of spinal implant assemblies intended for use in the occipito-cervical (OC) and cervico-thoracic (CT) regions (from the skull to the upper back). It simulates a worst-case scenario: a complete vertebrectomy (removal of a vertebra), creating a highly unstable spine segment that the implant must stabilize.

ASTM F2706 establishes standardized mechanical test methods to evaluate the static (strength) and fatigue (long-term durability) performance test, defines in‑vitro mechanical test protocols for posterior occipital‑cervical (OC) and occipital‑cervical‑thoracic (OCT) spinal fixation assemblies under a vertebrectomy worst‑case model, aiming for relative mechanical comparison of implant designs, rather than defining pass‑fail performance thresholds. 


ASTM F1264 Intramedullary Fixation Device Torsion & Bend Fatigue Test –

ASTM F1264 Standard Specification and Test Methods for Intramedullary Fixation Devices

ASTM F1264 is a standard specification used to describe the design and mechanical function of intramedullary fixation devices (IMFDs). It establishes terminology, classification, material, marking‑labeling requirements and four mandatory‑annex mechanical test methods for intramedullary nails/rods used in skeletal fracture fixation. It does not define mandatory clinical performance thresholds; tests deliver comparative mechanical data rather than absolute clinical pass‑fail criteria.

it also includes four standard testing methods: 

1, static four point bending test method (Annex A1), 

2, static torsion test method (Annex A2), 

3, bending fatigue test method (Annex A3), 

4, bending fatigue test method for IMFD locking screws (Annex A4). 


The UTDS series Dynamic/Fatigue Testing machine used for bone needle,  bending intramedullary nail fatigue test is mainly used for biomechanical performance testing of various materials, including artificial blood vessels, soft tissues, bones, bone plates, and vertebrae Interfusion device, knee joint, spine Dynamic and static mechanical tests on materials such as column fixators, metal coatings, hip joints, and intramedullary nails, including tension, compression, bending, and torsion.

ISO 7206-4, ISO 7206-6: Fatigue testing of femoral stem and femoral neck of hip prostheses –

ISO 7206‑4 and ISO 7206‑6 are key international standards defining fatigue endurance tests for stemmed femoral components of partial and total hip joint prostheses. ISO 7206‑4 evaluates the femoral stem under simulated proximal loosening; ISO 7206‑6 evaluates the femoral neck under well‑fixed, normal in‑vivo loading. 

ISO 7206-8 Hip Joint Replacement Torsion Testing, endurance properties of stemmed femoral components –

ISO 7206-8: Implants for surgery -- Partial and total hip joint prostheses -- Part 8: Endurance performance of stemmed femoral components with application of torsion. 


ISO 7206-8 standard as a series of test methods capable of effectively evaluating the endurance performance of stemmed femoral components designed for total and partial hip joint prosthesis. The standard allows experts to subject test specimens to loading conditions that mirror those seen in real conditions when patients use their hip joint prosthesis. The standard’s objective is to determine if hip implants perform as expected, ensuring their functionality and safety will be reliable for the patients who depend on them.


FAQS: ASTM F2345 Ceramic Modular Femoral Head Cyclic Fatigue Test

Q1. What is ASTM F2345?

ASTM F2345 is an ASTM International test method standard that describes how to evaluate the cyclic fatigue strength of ceramic modular femoral heads mounted on the metal cone (taper/trunnion) of a femoral stem used in total hip arthroplasty (THA/THR). The current edition is F2345-21. 


Q2. Is it a static test or a fatigue test?

It is a fatigue test method. Note the history: the previous edition (F2345-03(2013)) was titled "Static and Cyclic Fatigue Strength…". The 2021 revision dropped the standalone static method and now covers three methods: on-axis fatigue (7.4), off-axis fatigue (7.5), and a post-fatigue burst test (7.6) performed to ISO 7206-10 to measure the residual static strength of survivors.


Q3. How does F2345 relate to ISO 7206-10?

ISO 7206-10 is the companion static standard and is referenced twice inside F2345-21: for specimen characterization/assembly (clauses 6.1, 6.3, 7.2) and for the post-fatigue burst test (7.6). In short: ISO 7206-10 gives you static strength; F2345-21 adds the cyclic dimension and then hands the survivors back to ISO 7206-10 to measure what is left.


Q4. Which materials does it cover?

Primarily ceramic heads on metal cones: high-purity alumina (ASTM F603, ISO 6474-1), alumina-matrix composite with zirconia reinforcement (ISO 6474-2), and Y-TZP zirconia (ISO 13356). The standard states the methods "may have application to other materials."


Q5. Is ASTM F2345 recognized by FDA?

Yes. ASTM F2345-21 is listed in the FDA Recognized Consensus Standards database for medical devices (FR Recognition Number 11-385, date of entry 20 December 2021), with the extent of recognition given as "Complete standard."


Q6. Why is cyclic fatigue testing so important for ceramic femoral heads?

Ceramics such as alumina, ZTA and Y-TZP have no plastic deformation capacity and are flaw-controlled: their strength is governed by the largest defect population, and in a wet, 37 °C body environment those defects grow by subcritical (slow) crack growth. A single static burst number therefore says nothing about how strength decays over millions of walking cycles. F2345 reproduces that decay so engineers can see whether a design survives 10 million cycles and how much static strength is left afterwards.


Q7. What is the critical failure location, and why?

The conical bore of the head. The interference (friction-lock) fit between the metal cone and the ceramic bore converts the compressive joint load into tensile hoop stress in the bore — the worst possible loading mode for a brittle ceramic. Off-axis loading (L = 30°) further concentrates stress and reproduces the clinically observed head-fracture mode.


Q8. What clinical complications is this test guarding against?

The standard's rationale (Appendix X1.3) lists the potential complications of modular femoral heads: fracture of the femoral head; dislocation of the head from the cone; wear of the cone due to loosening/rotation of the head; and fracture of the cone. Ceramic head fracture is rare historically but catastrophic when it happens, which is why a discriminating screening test is required.


Q9. Can F2345 results predict in-vivo implant lifetime?

No — and the standard says so explicitly. Clause 4.2 states that in-vivo loading will generally differ from the loading defined in the methods and that the results "cannot be used to directly predict in-vivo performance." The method is a simplified comparative model (4.4), valid for ranking different head and cone designs and materials tested under similar conditions. Section 9 also states that no precision statement and no bias statement can be made, because there is no standard implant available to all users and no acceptable reference values.


Q10. Why is it important for material selection and QC specifically?

Because fatigue life is extremely sensitive to variables that are otherwise invisible in a static test: bore angle, bore major/minor diameter, cone angle and diameter, surface roughness (Ra, Rz per DIN 4768), circularity, length of the bore/cone mating interface, the installation (assembly) load, and the sterilization method. Clause 8.4 lists these as the optional data that make results interpretable. Two lots of the same ceramic with different bore finish can rank very differently under F2345.


Q13. Does the standard specify the peak load value?

No. The standard specifies the frequency, waveform, load ratio, alignment tolerance, medium and run-out, but the peak load magnitude is a user-selected test variable that must be reported. What the standard does cap is the equipment envelope: fatigue fixtures with capacity up to 50 kN, and static fixtures capable of sustaining forces up to the anticipated fracture level. Historically, static burst levels for this class of component exceed 40 kN, which is why a 50 kN-class frame is the practical minimum.


Q14. How many specimens are needed?

A minimum of five specimens is recommended per test group (7.4.3), and the report must include a statement justifying the number and kinds of samples (8.3). New cones and heads must be used for each test, and components that survive the test shall never be used for clinical purposes (7.3.2).


Q15. What is the post-fatigue burst test for, and when must it be run?

It quantifies how much strength the fatigue cycling actually removed. Survivors of the 10-million-cycle run are burst-tested statically per ISO 7206-10 in air at ambient conditions. Note 2 recommends doing it within a reasonable interval — for example one to two days — after fatigue loading ends, and the actual interval must be reported.


Q16. How long does one test take?

At the 30 Hz ceiling, 10 million cycles take about 92 hours (~3.9 days) per specimen; at 15 Hz roughly 7.7 days; at 5 Hz roughly 23 days. With a five-specimen group, a single-station machine can be occupied for weeks. This is why long-duration stability, unattended 24/7 operation and multi-station configurations matter so much for this standard.


Q17. Can one machine perform both the fatigue cycling and the burst test?

Yes — this is the normal configuration. A servo-hydraulic frame runs the sinusoidal fatigue program first and then, on the same load train, executes the quasi-static ramp to fracture per ISO 7206-10. The frame must simply be rated for the higher of the two demands (in practice the static burst level, not the fatigue amplitude).


Q18. Why choose ASTM F2345‑21 Cyclic Fatigue Test Machine for Ceramic Modular Femoral Heads from UnitedTest? 

UnitedTest supplies fully‑compliant ASTM F2345‑21 cyclic fatigue test machines for ceramic modular femoral heads of total hip replacement implants. Our servo‑hydraulic fatigue test system supports on‑axis, off‑axis fatigue and post‑fatigue burst test according to ASTM F2345‑21 & ISO 7206‑10 for orthopedic implant R&D, quality control and medical‑device regulatory validation.


UnitedTest is professional manufacturer of ASTM‑compliant orthopedic implant biomechanical testing machines. Our ASTM F2345‑21 ceramic modular femoral head cyclic fatigue tester is purpose‑built for evaluating cyclic fatigue strength of alumina, alumina‑zirconia composite and Y‑TZP zirconia femoral heads used in total hip arthroplasty (THR) implants.

Designed strictly following ASTM F2345‑21 requirements, UnitedTest servo‑hydraulic fatigue test system supports full test workflows including on‑axis fatigue test, 30° off‑axis physiological fatigue test, and post‑fatigue static burst test complying with ISO 7206‑10.


✅ Key machine capabilities & compliance

  • Max dynamic fatigue force up to 50 kN, meets ASTM E‑4 force‑verification standard; dynamic load control accuracy ± 2 % full‑scale.

  • Supports sinusoidal cyclic waveform; test frequency adjustable up to 30 Hz as per ASTM F2345‑21 limit.

  • Complete test fixture kit: hardened‑metal loading cone, replaceable copper‑ring holders, polymeric concave off‑axis loading insert, specimen assembly jigs.

  • Temperature‑controlled circulating bath for 37 °C physiological saline environment simulation.

  • Heavy‑duty safety protective shielding to contain flying ceramic fragments during brittle fracture events.

  • Software automatically records cycle count, detects fracture event, exports standardized test report templates for regulatory submission.

  • Compatible with all common femoral head sizes: 22 mm, 26 mm, 28 mm, 32 mm, 36 mm, 40 mm,44 mm.

Who uses UnitedTest ASTM F2345‑21 test machine?

  • Orthopedic hip implant manufacturers for new material screening, taper‑fit design iteration, manufacturing‑process validation.

  • Medical‑device test laboratories for third‑party mechanical performance verification.

  • University & research institute biomechanics labs for ceramic hip‑joint implant research.

  • Regulatory‑pre‑market submission mechanical performance testing for FDA‑recognized ASTM F2345‑21 consensus standard.

UnitedTest provides one‑stop solution: standard‑compliant hardware fixtures, installation, on‑site commissioning, operator training and long‑term after‑sales support for hip implant mechanical testing.

< Previous: ASTM F2118 Bone Cement Fatigue Test

> Next: ASTM F2346 Artificial Spinal Disc Static & Dynamic Fatigue Test

Require More Customized Solutions?

We offer customization to meet your specific needs. Our expert team will collaborate with you to develop the perfect product for you
Customize Now

Beijing United Test Co., Ltd.