ASTM F384 Testing for Metallic Angled Orthopedic Fracture Fixation Devices

Share:

ASTM F384 Tester for Metallic Angled Orthopedic Fracture Fixation Devices | UnitedTest

UnitedTest manufactures ASTM F384 compliant testing machines for metallic angled orthopedic fracture fixation implants. Perform single-cycle compression bend and bending fatigue tests to evaluate geometric and mechanical performance for surgical skeletal internal fixation devices.

ASTM F384 delivers complete specifications and standardized test protocols for metallic angled orthopedic fracture fixation devices used in skeletal internal fixation surgery. It defines consistent rules for classification, geometric definition and performance characterization, covering single-cycle compression bend testing and bending fatigue testing to assess mechanical properties relevant to in vivo clinical performance. UnitedTest supplies professional laboratory test equipment built to meet ASTM F384 requirements for orthopedic implant quality verification and regulatory testing.


Specific Test Methods

1, Single‑Cycle Compression‑Bend Test
Purpose

Evaluates static mechanical performance: measures compression‑bending stiffness (K, N/m) and compression‑bending strength (N·m) under combined compression and cantilever bending load. 

It uses a 0.2 % offset‑displacement criterion to determine the bending moment producing permanent plastic deformation. If fracture occurs before reaching proof load, fracture moment is taken as strength value.

Test principle

A static displacement‑controlled load is applied to the angled portion of the fixed device, generating combined compressive force and cantilever‑bending moment simulating primary in‑vivo loading for metaphyseal fracture implants.

  • Compression‑bending stiffness K: slope of linear‑elastic segment of load‑displacement curve.

  • 0.2 % offset displacement q = 0.002 × lever‑arm length L. Draw offset line parallel to elastic slope; intersection with load‑displacement curve yields proof load P. Compression‑bending strength = P × L. If fracture precedes intersection, strength = Fₘₐₓ × L (Fₘₐₓ = fracture load).

Test Equipment

1. Universal mechanical test machine complying with **ASTM E4** (force calibration and verification for testing machines).

2. Custom test fixture / anchor block to rigidly secure the sideplate of angled devices; alternative fixture configurations are permitted as long as loading requirements are satisfied.

3. Displacement acquisition system for measuring displacement of load application point, parallel to sideplate long‑axis.

4. Lever‑arm monitoring sensors, required if fixture or loading adapter translates perpendicular to machine loading axis during testing, for curve correction.

5. Data recording system to produce load‑versus‑displacement curves.

Machine‑fixture combined compliance shall be measured and reported; it should preferably be less than 1 % of specimen compliance.

ASTM F384 Testing for Metallic Angled Orthopedic Fracture Fixation Devices

Test Parameters
  1. Control mode: displacement‑controlled testing is strongly preferred; recommended test rate = 10 mm/min.

  2. Lever‑arm length: set to 80 % of blade‑length or longest lag‑screw length. Lever‑arm must be kept constant for comparative testing.

  3. Evaluation criterion: 0.2 % offset displacement for static bending strength.

Test Procedure

1. Prepare unused specimens; measure and document initial lever‑arm length.

2. Secure the angled‑device sideplate rigidly to test fixture, align load application point according to 80 % lever‑arm rule.

3. Set displacement‑controlled test rate to 10 mm/min; apply monotonically increasing displacement load.

4. Record continuous load‑versus‑displacement data until specimen fractures or sufficient post‑yield deformation is captured.

5. From recorded curve: fit the linear‑elastic slope Om; compute 0.2 % offset displacement q = 0.002 × L; draw offset parallel line BC.

6. Find intersection proof‑load P; compute compression‑bending strength = P · L; if fracture occurs before intersection, use fracture‑load Fₘₐₓ · L as strength.

7. Calculate compression‑bending stiffness K as slope of line Om.

8. Compile test report including specimen information, fixture configuration, lever‑arm, stiffness, strength statistics, fracture count, test‑control parameters.

2, Bending‑Fatigue Test
Purpose

Evaluates cyclic load resistance under sinusoidal cantilever bending. 

It generates M‑N (maximum moment‑to‑failure‑cycle) diagrams or estimates median fatigue strength at designated cycles (recommended at 1 000 000 cycles) for comparative assessment of cyclic durability.

Test principle

Sideplate is rigidly anchored. Sinusoidal cyclic bending moment is applied in cantilever‑bending configuration. 

Specimens are cyclically loaded until fracture, predefined termination limit, or run‑out cycle count (e.g., 1 000 000 cycles). 

Test outputs are moment magnitude and cycles‑to‑failure, for constructing M‑N curves or calculating median fatigue strength (50 %‑survival probability at target cycles) at fixed R‑ratio.

Test Equipment

1. Cyclic fatigue test machine satisfying ASTM E4 force‑verification requirements.

2. Cantilever‑bend test fixture compatible with Annex A1 fixture requirements.

3. Cycle counter to accumulate total loading cycles applied to specimen.

4. Limit‑detector system to detect specimen fracture / load deviation and terminate test automatically.

5. Data‑acquisition system evaluated according to **ASTM E1942**, to record maximum / minimum bending moment and cycle‑to‑failure data.

Test Parameters
  1. Waveform: sinusoidal cyclic bending‑moment waveform.

  2. Recommended R‑ratio (minimum moment / maximum moment) = 0.1. Deviations must be justified and reported.

  3. Suggested test frequency = 5 Hz, avoiding strain‑rate‑dependent material effects.

  4. Reference target for fatigue‑strength estimation: 1 × 10⁶ cycles run‑out.

  5. When setting test moment levels for exploratory testing: initial moments can correspond to 75 %, 50 %, 25 % of static compression‑bending strength measured in Annex A1. Moment‑level step size for staircase‑method fatigue‑strength testing shall not exceed 10 % of static bending strength.

Test Procedure

1. Select test moment levels; decide whether to produce M‑N diagram or to estimate fatigue strength at target cycles.

2. Mount specimen rigidly in cantilever‑bend fixture, configure lever‑arm consistent with Annex A1 loading geometry.

3. Configure cyclic test machine: sinusoidal moment waveform, set R‑ratio = 0.1, frequency = 5 Hz, cycle counter and failure‑limit detector.

4. Start cyclic loading, continue until specimen fracture, trigger‑limit activation, or run‑out cycle count is achieved.

5. Record for each specimen: maximum bending moment, total cycles‑to‑termination, failure location and failure mode.

6. For M‑N diagram: plot moment vs cycles‑to‑failure on semi‑log axes and apply curve‑fitting. For fatigue‑strength assessment, apply staircase / modified‑staircase statistical methods to estimate median fatigue strength at N cycles.


Related Referenced Standards

ISO 9585Implants for Surgery — Determination of Bending Strength and Stiffness of Bone Plates. 
ISO 14602Non‑active Surgical Implants — Implants for Osteosynthesis, Particular Requirements, referenced for implant marking requirements.
ASTM F382

Standard Specification and Test Method for Metallic Bone Plates

ASTM F384ASTM F384 Standard Specifications and Test Methods for Metallic Angled Orthopedic Fracture Fixation Devices
YY/T 1503Test method for fatigue performance of metal bone plates for surgical implants
YY/T 0119.4

Spinal implants.Components used in the surgical fixation of the spinal skeletal system.Part 4:Metallic spinal rods


Test Specimen Information

  1. Specimens must represent real implant production quality: identical material, cross‑section geometry, surface finish, marking and manufacturing processes; deviations must be documented in reports.

  2. Sample‑size rules:

    • Annex A1: Follow ASTM E122 for sample‑size calculation; minimum sample size = 3 when E122 data are unavailable.

    • Annex A2 M‑N diagram testing: minimum 3 replicates for one load level; statistically preferred setup: 5 replicates at 5 different moment levels.

    • Annex A2 fatigue‑strength estimation: about 20 specimens are typically needed for statistically meaningful results.


Industry Application Field

  • Orthopedic trauma implant manufacturers — design verification and validation, design-for-reliability, competitive benchmarking of blade plates, condylar/angled plates, and compression hip screw (sliding hip screw) systems.

  • Regulatory submissions — FDA 510(k), CE/MDR technical documentation, and China NMPA registration dossiers for metallic osteosynthesis implants; the standard supplies the recognized mechanical-characterization language for these filings.

  • Materials suppliers and forgers — qualifying F136/F138 bar, wire and forgings (F620/F621) in a finished-device context.

  • Test laboratories and CROs, and test-machine/fixture vendors who build angled-device compression-bend and fatigue rigs.

  • Clinical/academic biomechanics — comparative, rank-order evaluation of designs; explicitly not for predicting individual patient outcome


Keywords: ASTM F384 orthopedic implant tester, angled fracture fixation device testing machine, metallic orthopaedic fixation test equipment, compression bend test for orthopedic devices, bending fatigue tester for fracture implants, orthopedic surgical implant mechanical test, skeletal internal fixation lab tester, orthopaedic device quality control instrument, UnitedTest ASTM F384 tester, medical implant mechanical testing system

Related products and device

ASTM F384 Fatigue bending test machine for Metallic Angled Orthopedic Fracture Fixation Devices

Electronic Dynamic Universal Testing Machine/ Cyclic Testing Equipment widely used at material and components, parts dynamic mechanical property test, include tension, compression, low cycling test, match with high-low temperature chamber can execute the temperature test.

ASTM F384 metallic orthopaedic fixation test equipment

UnitedTest universal testing machine with various test fixture, like peel, flexure, puncture, tear, pneumatic grip, belt tension etc., can be used to almost all materials include but not limited to steel rod, rubber, steel wire, plastic, seat belt, textile, wood, panel etc., mechanical performance inspection.

orthopedic internal fixation implant bending tester

Electronic Fatigue Testing Machine designed for long-term cyclic load testing of metals, composites, and structural components, offering precise electronic control, repeatable results, and compliance with industrial standards.

ASTM F382 Multi‐Axis Fatigue Torsional & Bending testing system

Electronic Dynamic Testing Machine for high-precision cyclic and vibration testing of materials and components, featuring accurate frequency control, stable performance, and laboratory-grade reliability.

osteosynthesis bone plate regulatory compliance tester

Spinal Implant Constructs nail screw static and fatigue testing machine used to check the fatigue failure of medical nail, Loads are typically applied in a constant-amplitude, load-controlled sinusoidal waveform that runs more than 5 million times.

Related Standard

ISO 25539-3 Test methods of cardiovascular implants - Vena cava filters –

ISO 25539-3 defines mechanical performance tests as core bench evaluations for filter systems (filter + delivery/retrieval/conversion systems), plus supplementary tests for materials, imaging, biocompatibility, clinical validation, sterilization and packaging. Include the Filter Tensile Strength, Torsional Bond Strength,Tensile Bond Strength, Radial Force, Force to Deploy test etc.,

ISO 17137 Test Methods of cardiovascular absorbable implants –

Based on the ISO/TS 17137 standard the evaluation of mechanical properties is a critical part of the design and safety validation process. Mechanical Evaluation like tension, radial force, Cyclic Fatigue Durability like constant pulsing. Because absorbable implants lose their structural integrity over time through degradation, their mechanical performance must be assessed not just at the time of implantation, but throughout their intended functional lifetime. 

ISO 7198 Mechanical Tests and Other Evaluations for Vascular Implants –

ISO 7198 specifies test and performance requirements for tubular vascular grafts and vascular patches (surgical vascular prostheses). It defines mechanical tests (with full method/equipment/sample details) and other physicochemical, biological, dimensional, preclinical/clinical tests.

ASTM F2942 Vascular Implants Axial, bending, torsional and compression durability testing –

ASTM F2942 specifies in vitro test methodologies to evaluate the cyclic durability of vascular stents under non-radial mechanical deformations (axial, bending, and torsion) that occur in vivo due to musculoskeletal motion, breathing, or cardiac activity. 

ASTM F1798 Spinal Implant Subassembly Static and Fatigue Testing System –

ASTM F1798 Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants.

ASTM F1798 provides standardized methods for mechanically testing the interconnections within spinal implant systems. It is for evaluating the uniaxial static/fatigue strength and loosening resistance of interconnection mechanisms in spinal arthrodesis implant subassemblies, providing a standardized way to characterize mechanical performance of connections like rod-clamp, screw-rod, and hook-rod assemblies. It is critical for design validation, regulatory compliance, and clinical safety in the spinal implant industry.

ASTM F2502 Test of Internal Fixation Implants Bioabsorbable Plates and Screws –

ASTM F2502 Standard Specification and Test Methods for Absorbable Plates and Screws for Internal Fixation Implants

ASTM F2502 defines the requirements and mechanical test methods specifically for bioabsorbable (or biodegradable) orthopedic implants used to stabilize bone fractures. Unlike permanent metal implants, these devices are designed to provide temporary support and then gradually dissolve in the body. This standard ensures they possess sufficient initial mechanical strength and that their degradation properties are characterized in a standardized way. 

ASTM F2706 Fatigue Static Test of Spinal implant constructs –

ASTM F2706 is a critical biomechanical evaluation tool in the medical device industry, specifically for spinal implants. ASTM F2706 establishes standardized mechanical test methods to evaluate the static (strength) and fatigue (long-term durability) performance of spinal implant assemblies intended for use in the occipito-cervical (OC) and cervico-thoracic (CT) regions (from the skull to the upper back). It simulates a worst-case scenario: a complete vertebrectomy (removal of a vertebra), creating a highly unstable spine segment that the implant must stabilize.

ASTM F2706 establishes standardized mechanical test methods to evaluate the static (strength) and fatigue (long-term durability) performance test, defines in‑vitro mechanical test protocols for posterior occipital‑cervical (OC) and occipital‑cervical‑thoracic (OCT) spinal fixation assemblies under a vertebrectomy worst‑case model, aiming for relative mechanical comparison of implant designs, rather than defining pass‑fail performance thresholds. 


ASTM F1717 Spinal Implant Constructs Test –

ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Modelis a fundamental standard for evaluating the static and fatigue mechanical performance of spinal implant assemblies. Unlike ISO 12189, which focuses on a single device(an interbody cage), ASTM F1717 tests the entire posterior spinal construct—typically rods, screws, and cross-connectors—stabilizing a segment where a vertebral body has been removed (a "vertebrectomy" or "corpectomy" model). It simulates worst-case spinal instability to assess the implant's ability to maintain alignment and support loads.

ASTM F2068 Fatigue test system of Femoral Prostheses Implants, Hip Implants Testing –

ASTM F2068: Standard Specification for Femoral Prostheses – Metallic Implants


ASTM F2068 covers metallic stemmed femoral prostheses used to replace the natural hip joint by means of hemi-arthroplasty or total hip surgical procedures. Prostheses for hemi-arthroplasty are intended to articulate with the natural acetabulum of the patient. Prostheses for total hip replacement are intended to articulate with prosthetic acetabular cups. Prostheses may have integral femoral heads or cones designed to accept modular heads. 

The mechanical strength (fatigue test ISO 7206-4, ISO 7206-6, shear strength test ASTM F1044, tensile test ASTM F1147, abrasion resistance test ASTM F1978), corrosion resistance, and biocompatibility of the head portions of one-piece integral implants are covered by this specification.

ASTM F382 Metallic Bone Plates fatigue bending test –

ASTM F382 Standard Specification and Test Method for Metallic Bone Plates

ASTM F382 unified terminology, geometric classification, material requirements, marking & packaging rules, and mechanical test protocols for metallic bone plates used in surgical skeletal internal‑fixation. It define two mandatory test methods:  Single‑Cycle Bend Test and Bending Fatigue Test. Both adopt four‑point bending configuration. It is of great significance for ensuring the safety and reliability of metallic bone plates, which are widely used to align and fix bone segments in orthopedic reconstructive surgeries. 

FAQs for ASTM F384 (Metallic Angled Orthopedic Fracture Fixation Devices Test)

Q1. What does ASTM F384 cover?

It covers metallic angled orthopedic fracture fixation devices — implants with a component set at an angle to the long axis of the bone, used for metaphyseal fractures. The standard defines their geometry, classifies them, sets material/marking/handling requirements, and provides two mechanical test methods: single-cycle compression bend (Annex A1) and cantilever bending fatigue (Annex A2).


Q2. Which implant types are covered, and which are excluded?

Two classes: blade plates (angled component fixed relative to the sideplate) and compression hip screws (lag screw free to translate through a barrel). Excluded: intramedullary hip screw nails and any angled device without a sideplate (Note 1 of the standard).


Q3. What is the difference between Annex A1 and Annex A2?

Annex A1 is a static, single-cycle test that produces compression bending stiffness and 0.2 % offset compression bending strength. Annex A2 is a cyclic test that produces fatigue life at given moment levels or a median fatigue strength at N cycles, plotted as an M–N diagram. A1 is the capacity test; A2 is the durability test — and durability is the clinically decisive one.


Q4. Why is the ASTM F384 test important?

For five reasons: (1) it tests the finished device, not a material coupon — junction geometry, holes, barrel, threads and surface finish all matter; (2) the blade/barrel-to-sideplate junction is a stress concentrator and the real failure site; (3) fatigue, not static overload, is how these implants fail clinically; (4) stiffness governs load sharing with healing bone — too stiff causes stress shielding, too compliant allows fracture-gap motion and non-union; (5) it is the recognized mechanical language for FDA 510(k), CE/MDR and NMPA submissions.


Q5. How is the specimen set up?

The sideplate is rigidly anchored to a fully constrained anchor block. Load is applied parallel to the long axis of the sideplate, at a point on the angled portion, producing simultaneous compression and cantilever bending at the junction. Alternative setups (e.g. roller supports) are allowed if lever-arm change during the test is monitored and corrected for.


Q6. How is the lever arm determined?

The lever arm L is the distance from the line of load application to the bone-contacting surface of the sideplate at the most proximal fixture contact point. F384 requires the initial unloaded lever arm to be equivalent to 80 % of the blade length or of the longest screw, and it must be held constant across any comparative test series. Deviations must be noted and justified in the report.


Q7. What materials are permitted?

The device must be made from a metallic material intended for surgical implant use and biocompatible for the application. Common choices are ASTM F136 (wrought Ti-6Al-4V ELI, UNS R56401) and ASTM F138 (18Cr-14Ni-2.5Mo stainless, UNS S31673). Forged F136 must also meet F620; forged F138 must also meet F621.


Q8. How does ASTM F384 relate to ASTM F382?

Angled devices are formally a subset of bone plates, and F384 mandates F382 Annex A1 for the sideplate alone: bending stiffness, bending structural stiffness, bending strength and sideplate fatigue. So a full characterization uses F384 A1 + F384 A2 + F382 A1 — device-level compression bend, device-level fatigue, and four-point bend of the sideplate.


Q9. Can one UnitedTest machine run both Annex A1 and Annex A2?

In most cases yes, provided the frame, actuator, controller, sensors and test space are suitable and fatigue-rated fixtures are installed. Annex A1 needs a stiff static frame in displacement control; Annex A2 needs cyclic load control with sinusoidal waveform, cycle counting and limit detection. UnitedTest configures load capacity, frequency range, fixtures, gripping, environmental integration and data-acquisition software around your defined test method — send us the specimen drawing, standard revision, load range, amplitude, frequency, waveform, environment and target cycle count.


Q10. why choose ASTM F384 Testing Machine — Metallic Angled Orthopedic Fracture Fixation Device Test Solution from UnitedTest?

UnitedTest supplies fully‑compliant ASTM F384‑24 test systems for compression hip screw (DHS) and blade‑plate orthopedic implants. Perform Annex A1 single‑cycle compression‑bend test and Annex A2 bending‑fatigue test for static stiffness, bending strength and cyclic fatigue‑strength evaluation. Custom test fixtures, standardized reporting for orthopedic R&D and regulatory submission.


UnitedTest is a professional manufacturer of biomechanical testing equipment for orthopedic surgical implants. Our ASTM F384‑24 testing machine is purpose‑built to satisfy Annex A1 single‑cycle compression‑bend test and Annex A2 bending‑fatigue test requirements for metallic angled orthopedic fracture‑fixation devices, including compression hip screw (DHS) systems and blade‑plate implants.


Orthopedic implant manufacturers, biomechanics laboratories and medical‑device regulatory teams rely on ASTM F384‑24 bench tests to compare implant designs, evaluate titanium‑alloy and stainless‑steel implant performance, and generate mechanical data for pre‑market regulatory submissions (FDA‑recognized consensus standard).

Our complete ASTM F384‑24 test system includes: 

✅ Static‑dynamic universal test frame complying with ASTM E4 force‑verification standard 

✅ Custom‑designed compression‑bend cantilever fixture kit for DHS and blade‑plate specimens 

✅ High‑accuracy displacement and lever‑arm‑position monitoring sensors 

✅ Fatigue‑test modules: cycle counter, specimen‑failure limit detector, sinusoidal moment waveform generation 

✅ Dedicated test software: auto‑compute compression‑bending stiffness K, 0.2 % offset compression‑bending strength, plot M‑N fatigue diagrams, export full‑format ASTM‑compliant test‑reports

Our system supports core test parameters defined in ASTM F384‑24: displacement‑controlled static test rate 10 mm/min, configurable lever‑arm length, fatigue R‑ratio 0.1, cyclic frequency 5 Hz, 1 × 10⁶‑cycle run‑out setting for median‑fatigue‑strength estimation.

Application scenarios:

  • Orthopedic implant R&D: compare material grade, plate geometry, barrel‑junction structural designs

  • Quality‑verification for production batches of DHS / blade‑plate angled fracture‑fixation devices

  • Biomechanical laboratory bench‑testing

  • Mechanical‑performance documentation for medical‑device regulatory submissions

Note: ASTM F384‑24 test results are for relative performance comparison; they cannot directly predict absolute in‑vivo clinical performance of orthopedic implants.

UnitedTest provides one‑stop service: fixture customization, installation‑on‑site, equipment calibration, operation‑training and after‑sales technical support for your ASTM F384‑24 angled‑device compression‑bend and fatigue‑testing workflow. Contact UnitedTest to get your tailored ASTM F384‑24 testing‑machine quotation.

< Previous: ASTM F382 Metallic Bone Plates fatigue bending test

> Next: ASTM F543 Metallic Medical Bone Screws Test Driving Torque, Pullout Load Self-Tapping