Information on the most widely used ASTM standards within the materials testing industry
ISO 80369-20 Small-bore connectors for liquids and gases in healthcare applications – Part 20: Common test methods.
ISO 80369 standard outlines 9 specific test methods, each placed in an informative annex. These are broadly categorized into Leakage Tests, Mechanical Stress Tests, and Functional Tests. All tests use reference connectors defined in application‑specific parts (e.g., ISO 80369‑7) and uniform assembly rules.
Universal Preconditioning & Environment
Preconditioning: (20 ± 5) °C, (50 ± 10) % RH for ≥24h (required for hygroscopic materials; skip for non‑hygroscopic)
Test environment: 15–30 °C, 10–70 % RH (unless overridden by application parts)
Assembly:
Slip (non‑locking): Axial force 26.5–27.5N + torque ≤0.10Nm, rotation ≤90°, hold 5–6s
Locking: Torque 0.08–0.12Nm + axial force 26.5–27.5N, hold 5–6s
Details test stipulated in ISO 80369-20 introduction:
| 1, Leakage by Pressure Decay | |
| Test Method | Pressurize the connected assembly with dry air; judge leakage by pressure change (not leakage rate) |
| Stipulations | Defined start pressure, test volume, test period; max allowable pressure change set by application part. Absolute pressure change ≤ threshold defined by application part. |
| Samples | Representative production connectors (rigid/semi‑rigid); 60 samples per group for validation |
| Equipment | Pressure tester with pressure source, pressure gauge (±0.3% FS), timer, stop valve, test volume fixture, reference connector |
| Procedure | Seal sample; assemble to reference connector; Pressurize to specified value; close valve; Record start/end pressure and duration; Calculate pressure chang. |
| 2, Falling Drop Positive‑Pressure Liquid Leakage | |
| Test Method | Water (often dyed with methylene blue) is introduced into the connection and pressurized. |
| Stipulations | 300–330 kPa (per ISO 80369‑7), hold 30–35 s; horizontal orientation. No water droplets shall form or fall from the connection interface during the test. |
| Samples | Connector under test + appropriate reference connector (both dry before assembly). |
| Equipment | Water reservoir, pressurization system, pressure gauge (±0.3%), timing device. |
| Procedure | Assemble; fill with water, expel air Pressurize; hold duration Inspect for dripping |
| 3, Subatmospheric‑Pressure Air Leakage Test | |
| Test Method | Similar to Leakage by Pressure Decay, but a vacuum (negative pressure) is applied inside the connector instead of positive pressure. |
| Stipulations | 80.0–88.0 kPa vacuum, hold 15–20 s; max pressure change defined by application part |
| Equipment | Vacuum pump, pressure gauge, timer, test volume fixture |
| 4, Air Leakage During Aspiration | |
| Test Method | Apply vacuum while filled with water; detect continuous air bubbles |
| Stipulations | Specified subatmospheric pressure; wait 20 s for initial bubbles to stop |
| Equipment | Vacuum source, transparent water vessel, pressure gauge, timer |
| 5, Stress Cracking test | |
| Test Method | Evaluates if the connector material cracks when exposed to chemicals (e.g., disinfectants) under mechanical stress. |
| Stipulations | Assembly duration ≥48 h (extendable by application part) |
| Equipment | Fixture to maintain connection, leakage test setup |
| 6, Resistance to Separation from Axial Load | |
| Test Method | A direct pulling force (tension) is applied to the mated connectors. |
| Stipulations | The connectors must withstand a specified axial force (e.g., 23N to 35N depending on the application) for 10-15 seconds without separating. |
| Equipment | Tensile tester, force gauge, timer |
| Procedure | Assemble; apply axial force at ~10 N/s Hold; check sealing interface |
| 7, Resistance to Separation from Unscrewing | |
| Test Method | Applies to locking connectors (e.g., Luer Lock). A rotational unscrewing torque is applied. |
| Stipulations | Must withstand a specified torque (e.g., 0.018–0.020 N·m) for 10-15 seconds without disengaging. |
| Equipment | Torsion tester, timer |
| Procedure | Assemble locking connector; Apply unscrewing torque; hold; Inspect the connection condition. |
| 8, Resistance to Overriding | |
| Test Method | Tests the durability of the locking threads/lugs by applying a high overriding torque. |
| Stipulations | The locking mechanism must not strip, crack, or override when subjected to a specified maximum torque (e.g., 0.15–0.17 N·m). |
| Equipment | Torsion tester, timer |
| Procedure | Assemble; apply torque; Hold; inspect thread/lug engagement condition. |
| 9, Disconnection by Unscrewing Test | |
| Test Method | Ensures that connectors intended for daily disconnection (like syringe tips) can be easily and reliably unscrewed by the user without excessive force. |
| Stipulations | Peak torque ≤ limit set by application part |
| Equipment | Torsion tester with high sampling rate |
| Procedure | Assemble; rest 10–15 min; Increase unscrewing torque until separation; Record peak torque. |
Industry Field & Application
This standard is exclusively applied in the Healthcare and Medical Device Manufacturing industry. It governs the testing of small-bore connectors used in:
Intravenous (IV) therapy and hypodermic injections.
Respiratory ventilation and anesthesia.
Enteral feeding and urological drainage.
Neuraxial procedures (spinal/epidural).
Its ultimate goal is to provide a harmonized testing framework that guarantees the physical compatibility and safety of these life-critical interfaces.
Related Standards
ISO 18250 series: Connectors for medical reservoir delivery systems
ISO 14971:2019: Medical device risk management
IEC 62366‑1: Medical device usability engineering
IMDRF GRRP WG/N47: Essential safety/performance principles
EU 2017/745 MDR: European medical device regulation
Related products and device
Related Standard
ISO 80369 specifies the mechanical tests for verifying the safety and reliability of connectors under clinical conditions.The tests include:resistance to separation, axial pull-out force, air or liquid leakage resistance,stress cracking test etc.,
ISO 594-1 specified the requirements and test methods for syringes with a "slip" tip (a plain cone that relies on friction to hold the needle hub in place). ISO 594-2 Specifies locking-type (Luer lock) Luer connectors with thread/lug locking mechanisms.
ASTM F2477 designed to evaluate the long-term fatigue durability and radial cyclic deformation resistance of vascular implants under simulated physiological pulsatile loading conditions. It is crucial for simulating the cyclical stresses these medical devices endure inside human blood vessels.
ASTM F2942 specifies in vitro test methodologies to evaluate the cyclic durability of vascular stents under non-radial mechanical deformations (axial, bending, and torsion) that occur in vivo due to musculoskeletal motion, breathing, or cardiac activity. include Axial, bending, torsional, Pulsatile Durability, Radial Loading etc., test.
< Previous: ISO 80369 Test of Small-Bore Luer Connector for Liquids and Gases in Medical
> Next: None
Require More Customized Solutions?