Home >> Application >> By Standard >> ISO >> ISO 20000 + >> ISO 80369-20 Test methods Small-bore connectors for liquids and gases in healthcare

ISO 80369-20 Test methods Small-bore connectors for liquids and gases in healthcare

Share:

ISO 80369-20 Small-bore connectors for liquids and gases in healthcare applications – Part 20: Common test methods. 

ISO 80369 standard outlines 9 specific test methods, each placed in an informative annex. These are broadly categorized into Leakage Tests, Mechanical Stress Tests, and Functional Tests.  All tests use reference connectors defined in application‑specific parts (e.g., ISO 80369‑7) and uniform assembly rules.


Universal Preconditioning & Environment

Preconditioning: (20 ± 5) °C, (50 ± 10) % RH for ≥24h (required for hygroscopic materials; skip for non‑hygroscopic)

Test environment: 15–30 °C, 10–70 % RH (unless overridden by application parts)

Assembly:

Slip (non‑locking): Axial force 26.5–27.5N + torque ≤0.10Nm, rotation ≤90°, hold 5–6s

Locking: Torque 0.08–0.12Nm + axial force 26.5–27.5N, hold 5–6s


Details test stipulated in ISO 80369-20 introduction:

1, Leakage by Pressure Decay 
Test Method
Pressurize the connected assembly with dry air; judge leakage by pressure change (not leakage rate)
Stipulations

Defined start pressure, test volume, test period; max allowable pressure change set by application part.

Absolute pressure change ≤ threshold defined by application part.

SamplesRepresentative production connectors (rigid/semi‑rigid); 60 samples per group for validation
Equipment

Pressure tester with pressure source, pressure gauge (±0.3% FS), timer, stop valve, test volume fixture, reference connector

Procedure

Seal sample; assemble to reference connector;

Pressurize to specified value; close valve;

Record start/end pressure and duration;

Calculate pressure chang.

2, Falling Drop Positive‑Pressure Liquid Leakage
Test Method
Water (often dyed with methylene blue) is introduced into the connection and pressurized.
Stipulations

300–330 kPa (per ISO 80369‑7), hold 30–35 s; horizontal orientation.

No water droplets shall form or fall from the connection interface during the test.

SamplesConnector under test + appropriate reference connector (both dry before assembly).
EquipmentWater reservoir, pressurization system, pressure gauge (±0.3%), timing device.
Procedure

Assemble; fill with water, expel air

Pressurize; hold duration

Inspect for dripping

3, Subatmospheric‑Pressure Air Leakage Test
Test Method
Similar to Leakage by Pressure Decay, but a vacuum (negative pressure) is applied inside the connector instead of positive pressure.
Stipulations80.0–88.0 kPa vacuum, hold 15–20 s; max pressure change defined by application part
EquipmentVacuum pump, pressure gauge, timer, test volume fixture
4, Air Leakage During Aspiration
Test MethodApply vacuum while filled with water; detect continuous air bubbles
StipulationsSpecified subatmospheric pressure; wait 20 s for initial bubbles to stop
EquipmentVacuum source, transparent water vessel, pressure gauge, timer
5, Stress Cracking test
Test MethodEvaluates if the connector material cracks when exposed to chemicals (e.g., disinfectants) under mechanical stress.
StipulationsAssembly duration ≥48 h (extendable by application part)
EquipmentFixture to maintain connection, leakage test setup
6, Resistance to Separation from Axial Load
Test MethodA direct pulling force (tension) is applied to the mated connectors.
StipulationsThe connectors must withstand a specified axial force (e.g., 23N to 35N depending on the application) for 10-15 seconds without separating.
EquipmentTensile tester, force gauge, timer
ProcedureAssemble; apply axial force at ~10 N/s

Hold; check sealing interface

7, Resistance to Separation from Unscrewing
Test MethodApplies to locking connectors (e.g., Luer Lock). A rotational unscrewing torque is applied.
StipulationsMust withstand a specified torque (e.g., 0.018–0.020 N·m) for 10-15 seconds without disengaging.
EquipmentTorsion tester, timer
ProcedureAssemble locking connector; Apply unscrewing torque; hold; Inspect the connection condition. 
8, Resistance to Overriding
Test MethodTests the durability of the locking threads/lugs by applying a high overriding torque.
StipulationsThe locking mechanism must not strip, crack, or override when subjected to a specified maximum torque (e.g., 0.15–0.17 N·m).
EquipmentTorsion tester, timer
Procedure

Assemble; apply torque; Hold; inspect thread/lug engagement condition. 

9, Disconnection by Unscrewing Test
Test MethodEnsures that connectors intended for daily disconnection (like syringe tips) can be easily and reliably unscrewed by the user without excessive force.
StipulationsPeak torque ≤ limit set by application part
EquipmentTorsion tester with high sampling rate
ProcedureAssemble; rest 10–15 min; Increase unscrewing torque until separation; Record peak torque. 


Industry Field & Application

This standard is exclusively applied in the Healthcare and Medical Device Manufacturing industry. It governs the testing of small-bore connectors used in:

Intravenous (IV) therapy and hypodermic injections.

Respiratory ventilation and anesthesia.

Enteral feeding and urological drainage.

Neuraxial procedures (spinal/epidural).

Its ultimate goal is to provide a harmonized testing framework that guarantees the physical compatibility and safety of these life-critical interfaces.


Related Standards

ISO 80369 (1~7): General requirements for non‑interconnection & usability; Application‑specific connector dimensions (respiratory, enteral, neuraxial, intravascular, etc.

ISO 18250 series: Connectors for medical reservoir delivery systems

ISO 14971:2019: Medical device risk management

IEC 62366‑1: Medical device usability engineering

IMDRF GRRP WG/N47: Essential safety/performance principles

EU 2017/745 MDR: European medical device regulation

Related products and device

ISO 80369 medical connector Tensile strength testing machine

A single column tensile tester is a sophisticated material testing instrument designed for measuring mechanical properties of various materials under tension, compression, bending, shear, and other loading conditions. As the name suggests, it features a single vertical column supporting a moving crosshead that applies force to test specimens.

ISO 80369 Medical Luer Connector Multifunctional Testing Machine

Medical Luer Connector Multifunctional Testing Machine can detect the leakage, air leakage, separation force, unscrewing torque, silky resistance, overload resistance, stress cracking, pressure attenuation leakage, positive pressure air leakage, negative pressure air leakage of Luer joint.

ISO 80369 Luer Conical Joint Comprehensive Tester

Medical Luer Connector Multifunctional Testing Machine can detect the leakage, air leakage, separation force, unscrewing torque, silky resistance, overload resistance, stress cracking, pressure attenuation leakage, positive pressure air leakage, negative pressure air leakage of Luer joint.

ISO 80369 Medical infusion device flow volume tester

​Medical Device Flow Volume Tester is a specialized equipment designed to detect the flow of medical devices. Suitable for infusion sets, blood transfusion sets, infusion needles, catheters, anesthesia filters, etc.,

ISO 80369 Connector Torsion Tester

Manual torsion tester is mainly used to execute torsion test for various kinds materials, manual loading, high accuracy torsion transducer to measure torque, photoelectric encoder to measure torsion angle, digital display test result.

Related Standard

ISO 80369 Test of Small-Bore Luer Connector for Liquids and Gases in Medical

ISO 80369 specifies the mechanical tests for verifying the safety and reliability of connectors under clinical conditions.The tests include:resistance to separation, axial pull-out force, air or liquid leakage resistance,stress cracking test etc.,

ISO 594 Luer conical fittings test requirement

ISO 594-1 specified the requirements and test methods for syringes with a "slip" tip (a plain cone that relies on friction to hold the needle hub in place).  ISO 594-2 Specifies locking-type (Luer lock) Luer connectors with thread/lug locking mechanisms.

ASTM F2477 Test for in-vitro Pulsatile Durability Testing of Vascular Stents

ASTM F2477 designed to evaluate the long-term fatigue durability and radial cyclic deformation resistance of vascular implants under simulated physiological pulsatile loading conditions. It is crucial for simulating the cyclical stresses these medical devices endure inside human blood vessels. 

ASTM F3036 Testing of Absorbable Stents

ASTM F2942 specifies in vitro test methodologies to evaluate the cyclic durability of vascular stents under non-radial mechanical deformations (axial, bending, and torsion) that occur in vivo due to musculoskeletal motion, breathing, or cardiac activity. include Axial, bending, torsional, Pulsatile Durability, Radial Loading etc., test. 

< Previous: ISO 80369 Test of Small-Bore Luer Connector for Liquids and Gases in Medical

> Next: None

Require More Customized Solutions?

We offer customization to meet your specific needs. Our expert team will collaborate with you to develop the perfect product for you
Customize Now

Beijing United Test Co., Ltd.