Your current locationHome>Products>Medical/Biomedical device>Spinal implant, bone plates
Products
ISO 16402,ASTM F2118 Bone Cement Four-Point Bending Fatigue Tester

ISO 16402, ASTM F2118 Bone Cement Four-Point Bending Fatigue Tester


Main Applications

The Bone Cement Four-Point Bending Fatigue Tester is primarily used to evaluate the fatigue performance of bone cement materials under four-point bending conditions. Specifically, it can conduct mechanical performance tests, such as fatigue durability and lifespan, on specimens of bone cement, biomaterials, surgical implant materials, and medical materials. By simulating real-world stress conditions, the tester assesses the bending strength, fatigue life, deformation characteristics, and performance of bone cement under various environmental conditions. These test results are crucial for ensuring the reliability and safety of bone cement in practical applications, aiding in product design optimization, production process improvement, and overall product quality enhancement. 


Sample preapare: A four-point bending test was made on bone cements. The samples were cyclically loaded at a stress level of 20 MPa at a frequency of 4 Hz. The different cements were mixed either by hand or mechanically with or without a 0.15-bar vacuum. Mechanical mixing in a vacuum improved all three cements, and was the best of all the tested mixing techniques. Simplex P showed the best fatigue properties regardless of mixing technique.


Standard: ISO 16402:2008

Implants for surgery — Acrylic resin cement — Flexural fatigue testing of acrylic resin cements used in orthopaedics

ASTM F2118-14

Standard Test Method for Constant Amplitude of Force Controlled Fatigue Testing of Acrylic Bone Cement Materials

The ISO 16402, ASTM F2118 standard specifies the test procedures for determining the fatigue characteristics of resin-based bone cement made from polymethyl methacrylate, employing rectangular bar samples and conducting four-point bending tests under quasi-static or cyclic loading conditions to obtain the bending strength or fatigue life of the samples.We provide professional testing jigs (in a water bath environment) and precise and effective testing solutions (SOP) to assist in the analysis of the mechanical properties of bone cement.


Main Parameter of the UnitedTest fatigue test machine:

Max. dynamic load:±10KN  Accuracy ±1%

Max. static load: ±10kN

Dynamic fluctuation: Less than ±1%FS

Actuator max. stroke: ±50mm, accuracy ±0.5%FS

Amplitude at Max. frequency: ±1mm

Max. Frequency: 0.1HZ~15Hz

Max. line velocity: 150mm/s


For more information or details, please contact us:

Email: export@unitedtest.com 

whatsapp: +8613911165373


Max.   dynamic load

±10KN  Accuracy ±1%

Max.   static load

±10kN

Dynamic   fluctuation

Less   than ±1%FS

Actuator   max. stroke

±50mm,   accuracy ±0.5%FS

Amplitude   at Max. frequency

±1mm

Max.   Frequency

0.1HZ~15Hz

Max.   line velocity

150mm/s

Control   mode

Load,   displacement, deformation

Main   test waveform

Sinusoidal   wave, triangular wave, square wave, sawtooth wave etc.,

Pass   through Zero

Complete   display the pass through zero hysteresis loop curve.

Columns   number

2

Vertical   space

750mm

Max.   test space

450mm   (include test fixture)

Test   width

450mm

Cooling   method

Water   cooling

Weight

350kg

Dimensions

785*620*2050mm

Standard   Power

380/220V,   50/60HZ, 3 phase, 4Kw

Working   system

MS   Win7 / Win10

Load   cell

USA,   Vishay Celtron


Standard: ISO 16402:2008

Implants for surgery — Acrylic resin cement — Flexural fatigue testing of acrylic resin cements used in orthopaedics

The ISO 16402 standard specifies the test procedures for determining the fatigue characteristics of resin-based bone cement made from polymethyl methacrylate, employing rectangular bar samples and conducting four-point bending tests under quasi-static or cyclic loading conditions to obtain the bending strength or fatigue life of the samples.We provide professional testing jigs (in a water bath environment) and precise and effective testing solutions (SOP) to assist in the analysis of the mechanical properties of bone cement.

ASTM F2118-14(2020) 

Standard Test Method for Constant Amplitude of Force Controlled Fatigue Testing of Acrylic Bone Cement Materials


Your email address
Your message

Related products
Contact

Email:
export@unitedtest.com
unitedtest@hotmail.com
Tel:
+86-10-57055071
+86-13911165373

MS Teams: 
unitedtest@hotmail.com

WhatsApp:
+86 13911165373

WeChat:
UnitedTest-V1 

X