Information on the most widely used ASTM standards within the materials testing industry
Model: XJU-5.5/22, Dial scale, ISO 180 ASTM D256
Model XJU-5.5, XJU-22 Dial IZOD Impact Tester has a wide measurement range, a clear double-layer dial, is easy to operate, and provides accurate and reliable measurement results. This impact testing equipments is mainly used to test the impact toughness of non-metallic materials such as hard plastics, reinforced nylon, fiberglass, ceramics, stoneware, and electrical insulating materials. It is an ideal testing device for chemical industry enterprises, research institutions, universities and colleges, quality inspection departments, and specialized manufacturers' laboratories.
Instrument Structure
It uses a dual-energy pendulum structure design, achieving graded adjustment of impact energy through a rotating shaft fixing device. The main frame consists of three parts: the energy release mechanism, the specimen fixture, and the data acquisition system.
General Introduction
Model XJU-5.5, XJU-22 Dial IZOD Impact Tester has a wide measurement range, a clear double-layer dial, is easy to operate, and provides accurate and reliable measurement results. This impact testing equipments is mainly used to test the impact toughness of non-metallic materials such as hard plastics, reinforced nylon, fiberglass, ceramics, stoneware, and electrical insulating materials. It is an ideal testing device for chemical industry enterprises, research institutions, universities and colleges, quality inspection departments, and specialized manufacturers' laboratories.
Instrument Structure
It uses a dual-energy pendulum structure design, achieving graded adjustment of impact energy through a rotating shaft fixing device. The main frame consists of three parts: the energy release mechanism, the specimen fixture, and the data acquisition system.
Key Features
1, Dial display, Izod Test model;
2, This economical model is simple in structure, easy to operate.
3, Support vice, pendulum is nickel coating, good appearance and anti-corrosion.
4, Standard ISO 180, ASTM D256
5, Two energy range 1-5.5J, 5.5 ~ 22J for selection.
ISO 180: “Plastics -- Determination of Izod impact strength”
ASTM D256: “Standard Test Methods for Determining the Izod Pendulum Impact Resistance of Plastics”
Main Technical Specification
| Model | XJU-5.5 | XJU-22 |
| Impact energy | 1J, 2.75, 5.5J | 5.5J, 11J, 22J |
| Impact speed | 3.5m/s | 3.5m/s |
| Pendulum center to impact blade distance | 322mm | 322mm |
| Distance from impact blade to upper surface | 22mm | 22mm |
| Blade angle radius | R=0.8mm | R=0.8mm |
| Pendulum initial angle | 160° | 160° |
| Display type | Dial analog display | |
| Specimen clamping type | IZOD | |
| Standard Accessories | Loadframe, pendulum(1J, 5.5J), pendulum weight (2.75J), vice adjusting plate, spanner, manual etc. | Loadframe, pendulum(5.5J, 11J, 22J), vice, impact standard plate, spanner, manual etc. |
Main Accessories
| Standard Accessories | Loadframe, pendulum(1J, 5.5J), pendulum weight (2.75J), vice adjusting plate, spanner, manual etc. | Loadframe, pendulum(5.5J, 11J, 22J), vice, impact standard plate, spanner, manual etc. |