Information on the most widely used ASTM standards within the materials testing industry
Model: UT2054, IEC60068-2
Constant temperature and humidity test chamber Equipment for testing the performance of materials in various environments and testing various materials heat-resistant, cold-resistant, dry-resistant, moisture resistance. Suitable for electronics, electrical appliances, mobile phones, communications, instruments, vehicles, plastic products, metals, food, chemical, building materials, medical, aerospace and other products for quality testing.
It is composed of three parts: heating, humidification and refrigeration. By means of a circulating motor mounted on the top of the box, the air is discharged into the box to achieve gas circulation and balance the temperature and humidity in the box. The data collected by the humidity sensor is transmitted to the temperature and humidity controller (microprocessor) for editing Processing, issuing control instructions.
General Introduction
Constant temperature and humidity test chamber Equipment for testing the performance of materials in various environments and testing various materials heat-resistant, cold-resistant, dry-resistant, moisture resistance. Suitable for electronics, electrical appliances, mobile phones, communications, instruments, vehicles, plastic products, metals, food, chemical, building materials, medical, aerospace and other products for quality testing.
It is composed of three parts: heating, humidification and refrigeration. By means of a circulating motor mounted on the top of the box, the air is discharged into the box to achieve gas circulation and balance the temperature and humidity in the box. The data collected by the humidity sensor is transmitted to the temperature and humidity controller (microprocessor) for editing Processing, issuing control instructions.
Key Features
This test equipment prohibits:
Testing or storage of flammable, explosive and volatile substance specimens
Testing or storage of corrosive substance specimens
Testing or storage of biological specimens
Testing or storage of specimens of strong electromagnetic emission sources
Testing or storage of specimens of radioactive materials
Testing or storage of highly toxic substance specimens
Testing or storage of specimens that may produce flammable, explosive, volatile, highly toxic, corrosive and radioactive substances during testing or storage
| Volume | 408L |
| Inner size | W 800mm × D 600mm × H 850mm |
| Machine dimension | W 1350mm × D 1150mm × H 1950mm |
| Weight | 300KG |
| Power | 380V, 50HZ, 8KW |
| Insulated enclosure |
|
| Air conditioning channel A |
|
| Test chamber standard configuration |
(The cumulative total load of samples in the box does not exceed: 100kg) |
| Door | Single-open hinged door (left hinge, right handle) with observation window, lighting, window frame/door frame anti-condensation electric heating device. |
| Control panel | Controller display, power switch, and status indicator |
| Mechanical room | Refrigeration unit, water connection plate, drainage holeCondensing fan, condenser inlet, replenishment tank |
| Distribution control cabinet | IO boards, machine transformers, ballasts, intermediate relays, time relays, solid state relays, AC contactors, thermal relays, fuses, air switches |
| Humidifier(Humid heat type only) |
|
| waterway(Humid heat type only) |
|
| Power cord holes and drain holes | Located on the back of the cabinet |
| Mode of work A | Two levels compression |
| Refrigeration compressors | Fully enclosed low noise rotary lobe compressor |
| evaporator | Finned tubular heat exchanger (doubles as a dehumidifier) |
| condenser | Air-cooled: Finned tube heat exchanger. |
| Throttle device | Expansion valve + capillary |
| Plate heat exchanger(Cascade refrigeration system adopted) | Stainless steel brazed plate heat exchanger |
| Refrigeration control mode |
|
| Refrigerant | R404a/R23 (Ozone depletion index is 0, R23 is used for cascading refrigeration systems) |
| Monitor | 5 inch |
| Running way | Program mode, valuation mode |
| Setting way | Chinese and English menu (can be freely selected), touch screen input |
| Program capacity |
|
| Set the range | Temperature: Adjusts according to the temperature operating range of the device. |
| resolution | Temperature: 0.1°C; Time: 1min; Humidity: 0.1%RH (temperature and humidity test equipment) |
| Input | PT100 Platinum thermal resistance |
| Control mode |
|
| Curve recording function | Ram with battery protection to save the setpoint, sampling value and time of the sampling moment of the device; Maximum recording time is 180 days (when the sampling period is 60S) |
| Satellite features |
|
| Temperature and humidity measurement. | Temperature: PT100 Platinum thermal resistanceHumidity: Wet and dry bulb thermometer method (damp heat type only) |
GB/T2423.1-2008 (IEC60068-2-1:2007) Low temperature test method Ab
GB/T2423.2-2008 (IEC60068-2-2:2007) High temperature test method Bb
GJB150.4-1986 Low temperature test
GJB150.3-1986 high temperature test
GB/T2423.3-2006 (IEC60068-2-78:2007) Constant humid heat test method
GB/T2423.4-2008 (IEC60068-2-30:2005) Alternating damp heat test method
GJB150.9-1986 Damp heat test (Figure 1, Figure 2)
Main Technical Specification
| Test the environment conditions | The ambient temperature is +5 to +28 °C, the relative humidity is ≤ 85%, and there is no specimen in the test chamber |
| Test meth | GB/T 5170.2-2008 Temperature test equipment; GB/T 5170.5-2008 Damp heat test equipment |
| Temperature range | 0℃~+150℃ |
| Humidity range | 20%---98% Humidity Temperature (25 to 85 degrees Celsius can do humidity) |
| Temperature fluctuations | ≤0.5℃(≤±0.25℃) |
| Temperature deviation | ±2℃ |
| Temperature uniformity | ≤±2℃ |
| Temperature rising time | 3℃/min |
| Temperature falling time | 1℃/min |
| Humidity range(Humid heat type only) | (20~98)%RH
Temperature and humidity can be controlled range diagram |
| Relative humidity deviation(Humid heat type only) | ±2.0%RH |
| Meet the test method | GB/T2423.1-2008 (IEC60068-2-1:2007) Low temperature test method GB/T2423.2-2008 (IEC60068-2-2:2007) High temperature test method GJB150.4-1986 Low temperature test GJB150.3-1986 high temperature test GB/T2423.3-2006 (IEC60068-2-78:2007) Constant humid heat test method GB/T2423.4-2008 (IEC60068-2-30:2005) Alternating damp heat test method GJB150.9-1986 Damp heat test (Figure 1, Figure 2) (The load per cubic meter is not more than 35kg/m3 of steel heat capacity, and there is no active humidity or heat load during the damp heat test) |
| Hot and humid system water supply (hot and humid type only) | |
| Water supply mode | Pump lifting |
| Location of the water supply unit | Front drawer tank |
| Storage tank volume | 50L(Portable) |
| Water quality requirements for water supply | Resistivity ≥500Ω·m |
| Safety protection devices | |
| Refrigeration system |
|
| Humidification system(Humid heat type only) |
|
| Test chamber |
|
| Other |
|
| Other configurations | |
| Power cable | 380V 8KW |
| Total power switch | NFB breaker |
| Installation site |
|
A: Not less than 60cm B: not less than 120cm C: not less than 60cm D: not less than 120cm | |
| Environmental Conditions |
|
Tap water (Only for humid heat type equipment with pure water machine) | Flow rate ≥ 200kg/h, pressure 0.1MPa ~ 0.25MPa Tap water meets GB 5749-1985 drinking water sanitation standards Note: Pure water machine with DN15 female thread joint |