Home >> Products >> Impact Testing Machine >> Auxiliary Device >> Impact Test Sample UV Notch Profile Projector
Impact Test Sample UV Notch Profile Projector

Impact Test Sample UV Notch Profile Projector

Model: CST-50

CST-50 projector is a supplemental equipment for impact test, mainly used to check the accuracy of impact specimen U/V notch. Put the notched specimen on projector working table, multiple 50* and compare the projection image with standard plate, then identify the quality of impact specimen notch.


The impact specimen UV notch projector is a special testing equipment for metal Charpy impact test, which accurately measures the size and shape of the V-shaped (2mm) and U-shaped (2/3/5mm) notches of the impact specimen through the principle of optical amplification, meeting the requirements of GB/T 229, ISO 148 and other standards, and is a key quality inspection equipment to ensure the qualification of the specimen and the accuracy of the test data.


General introduction

CST-50 projector is a supplemental equipment for impact test, mainly used to check the accuracy of impact specimen U/V notch. Put the notched specimen on projector working table, multiple 50* and compare the projection image with standard plate, then identify the quality of impact specimen notch.


The impact specimen UV notch projector is a special testing equipment for metal Charpy impact test, which accurately measures the size and shape of the V-shaped (2mm) and U-shaped (2/3/5mm) notches of the impact specimen through the principle of optical amplification, meeting the requirements of GB/T 229, ISO 148 and other standards, and is a key quality inspection equipment to ensure the qualification of the specimen and the accuracy of the test data.

Key Features

  • Key Features
  • Core principle
  • Core Function
  • Standard
  • Use and Maintenance
  • Selection Recommendations

Magnified projection: Magnifies tiny notches by 10–50 times to clearly visualize the outline of the notch and is directly visible to the naked eye.

Precise Measurement: Measures critical dimensions such as notch angle (V-shaped 45°), depth (V-shaped 2mm, U-shaped 2/3/5mm), and bottom radius (U-shaped 1/1.5/2.5mm).

Intuitive judgment: Through the comparison of standard templates on the projection screen, it can quickly determine whether the notch meets the standard, and if it is not qualified, it can be reworked in time.

Digital display assistance: Some models have digital display function, automatic reading, reduce manual error, and improve detection efficiency.

Accuracy measuring: 

Impact Test Sample UV Notch Profile Projector

Principle: The specimen notch is enlarged (common 10×/20×/50×) and projected onto the projection screen, and the size is read by the ruler or digital display system, and the notch angle, depth, radius and other parameters are intuitively detected.


Structure: optical system (light source, objective lens, mirror, projection screen), table (pan/rotate with positioning fixture), measuring system (ruler / digital display), focusing mechanism, base (anti-vibration design).


Detect the dimensional accuracy and shape consistency of the notch of the impact specimen, exclude unqualified specimens, and avoid affecting the impact test results.


It is used in physical and chemical laboratories in metallurgy, steel, boiler pressure vessels, aerospace, machinery manufacturing, scientific research institutes and other fields.


Specimens prepared by manual / electric broaching machines are a necessary quality inspection link before impact testing to ensure that the test data is true and reliable.


GB/T 229-2007/2010, 

ASTM E23 “Standard Test Methods for Notched Bar Impact Testing of Metallic Materials”

ISO 148 “Metallic materials -- Charpy pendulum impact test” 

EN 10045.

Operation Process: 

1. Preparation: Place the equipment on a horizontal, shockproof countertop and reliably ground; Turn on the power and warm up for 5–10 minutes to stabilize the light source.

2. Loading: Place 10×10×55mm specimens into the table fixture, positioning and aligning, ensuring that the notch is facing the optical system, clamping and secure.

3. Focus and alignment: Adjust the magnification of the objective lens (usually 20×), turn the focus knob to make the notch outline on the projection screen clear; Move the table so that the center of the notch is aligned with the projection screen reticle.

4. Measurement and comparison: Measure the notch angle, depth, and bottom radius through the projection screen ruler or digital display system; Compare with the standard sample outline to determine whether it is qualified.

5. Recording and pick-up: record the measurement data, mark the qualified specimens and set them aside, and re-broach the unqualified specimens; Loosen the fixture, remove the specimen, and clean the workbench.

6. Power off: Turn off the power, cover the dust cover, and clean the dust on the surface of the projection screen.


Maintenance

weekly: use lens paper to clean optical components such as objectives and mirrors; Check the lubrication of the workbench guide rail, and apply special lubricating oil if necessary; Calibration of rulers or digital display accuracy.

Monthly: fasten the connecting bolts of the equipment to prevent loosening; Check whether the power plug and terminal blocks are in good contact; Clean the dust inside the equipment (after power outage).

Quarterly: Full calibration of measurement accuracy (standard gauge can be used); Check the flatness of the projection screen and adjust it if necessary; Replace the aging light source bulb or fuse.

Long-term deactivation: clean the optical parts and body, cover the dust cover after drying; Remove the light source bulb and store it separately; Regularly turn on the machine to warm up for 10 minutes and take care of the optical system.


Magnification: 10×/20× double gear for conventional detection; The radius at the bottom of the notch needs to be carefully measured, and three levels of 10×/20×/50× should be selected.

Measurement method: Pursue efficient selection of digital display (automatic reading); Budget limited selection manual ruler type (intuitive and convenient).

Specimen type: only the V-shaped notch is selected for special type; Taking into account the V/U-type notch, the general type can be selected (the standard model can be switched).

Laboratory environment: For environments with large vibrations, choose models with anti-vibration bases; Space is limited, and miniaturized design models are selected.


Main technical specification

ModelCST-50
Projector screen   diameter180mm
Magnifying   multiplication50*
Angle measuring   accuracy
Light source12V 100W
Working table moving360º
Stroke of working tableLongitudinal: ±10mm; Transverse: ±10mm
Size of working tableSquare working table diameter: 90mm
  Glass diameter: 70mm
Power220V 50HZ 200W/110V, 60HZ
086dhDimensions(mm)515×224×603
Weight18kg


Main accessories

Frame1   set
Power   cord1   pc
Halogen lamp2   pcs
Plane   glass plate1   pc
Documents   (Manual, packing list, certificate)


Safety points

Safety points

Avoid looking directly at the light source during operation to prevent strong light from irritating the eyes; It is forbidden to touch optical components such as objectives and mirrors during the operation of the equipment.

The specimen should be clamped firmly to prevent it from shifting during measurement; Move the table gently to avoid collision with the optics.

It is forbidden to wipe the projection screen and lens with hard objects, and it needs to be cleaned with special lens paper; Avoid placing strong magnetic fields and heat sources around the equipment to prevent affecting accuracy.

Regularly check the insulation of the power cord to avoid leakage; If there is abnormal noise or blurred image, stop the machine immediately for inspection.


Frequently Asked Questions and Countermeasures

issuecausecountermeasure
Blurry imagesDirty optics, unfocused objectives, and   aging bulbsClean the lens, refocus, replace the bulb
Large measurement errorThe ruler/digital display is not   calibrated, the specimen is offset, and the table is looseCalibrate equipment, re-clamp specimens,   fasten table bolts
The light source flickersThe power supply voltage is unstable, the   bulb is poorly contacted, and the fuse is damagedStabilize the power supply, reinstall the   bulb, and replace the fuse
There is a black shadow on the projection   screenThe position of the mirror is offset, the   angle of the light source is improper, and the specimen is occludedAdjust the mirror position, calibrate the   angle of the light source, and reposition the specimen


Require More Customized Solutions?

We offer customization to meet your specific needs. Our expert team will collaborate with you to develop the perfect product for you
Customize Now